Centrifuge speed: converting RPM to relative centrifugal force and back
A centrifuge spins samples so that the particles in them settle faster than gravity alone would make them. How hard they are pushed is the relative centrifugal force, RCF, written in multiples of Earth’s gravity, × g. Machines are set in revolutions per minute, though, and published methods give × g, so a conversion is needed that depends on the rotor. This calculator converts both ways for any radius and lists the force at a series of common speeds for your rotor.
The formula
A particle at distance r from the axis of a rotor spinning at angular speed ω feels a centripetal acceleration ω²r. Dividing by the acceleration of gravity gives the RCF: RCF = ω²r / g, with ω = 2π × RPM / 60 in rad/s, r in meters and g = 9.80665 m/s². With r in centimeters and RPM in revolutions per minute this reduces to RCF = 1.118 × 10⁻⁵ × r × RPM². The force rises with the square of the speed and in direct proportion to the radius.
Using the converter
Choose RPM to RCF or RCF to RPM, type the speed or the force and the Rotor radius with its unit: centimeters, millimeters or inches. The page opens on 10,000 RPM at 8.4 cm and shows 9,393.2492 × g, the radius in three units, the formula and the angular speed 1,047.1976 rad/s. Sample chips load a conversion from 12,000 × g, a gentle cell spin of 300 × g with a 15 cm radius and a rotor of 5.5 inches.
Worked examples
To reach 12,000 × g with an 8.4 cm radius the centrifuge must run at 11,302.713 RPM, with an angular speed of 1,183.6173 rad/s. A cell-pelleting spin of 300 × g in a rotor with a 15 cm radius needs 1,337.355 RPM. A 5.5 inch rotor, which is 13.97 cm, spinning at 3,000 RPM produces 1,405.9681 × g. Squaring explains why doubling the speed gives four times the force: 5,000 RPM at 8.4 cm is 2,348.31 × g and 10,000 RPM gives 9,393.25.
The table
A table lists the force at 1,000, 2,000, 3,000, 4,000, 5,000, 6,000, 8,000, 10,000, 12,000, 14,000, 15,000 and 20,000 RPM for the radius you entered: at 8.4 cm that is 93.9325, 375.73, 845.392, 1,502.92, 2,348.31 and 3,381.57 × g for the first six speeds. It gives a feel for the range of your rotor and a quick way to choose a speed close to a protocol’s value.
Choosing the radius
The force depends on the distance from the axis, so one tube sees a range of forces from the top of the liquid to the bottom. Manuals quote the minimum, average and maximum radii of a rotor, and the maximum, at the tube bottom, gives the highest force and is the figure most protocols assume. Fixed-angle and swing-bucket rotors have different radii, so a conversion made for one does not apply to the other, and a different tube size changes the radius.
Practical points
Do not exceed the maximum speed and the maximum load of the rotor and tubes. Balance the tubes. Convert a protocol to the speed of your own machine rather than copying RPM from another lab, and check whether your centrifuge displays RPM or RCF directly. Setting a speed with the nearest available step is usually fine; the force will differ from the target by a few per cent, which rarely matters for pelleting but may for density gradients.
Limits
The calculation uses a single radius and the ideal formula. It does not account for the acceleration phase, for the change in radius as a tube swings out, for temperature or for the rotor’s manual limits, and it cannot say which speed is right for your sample. Follow your laboratory’s procedures and the manufacturer’s instructions for safe operation. The values are arithmetic aids.
Related tools
- Hz to RPM converts rotation speeds between hertz, RPM and rad/s.
- Serial Dilution Calculator plans the dilutions that often precede a spin.
- Primer Tm Calculator is another everyday molecular biology calculation.
- Energy Converter converts the energy units that appear in related physics.