The Beer-Lambert law: absorbance, concentration, path length and transmittance
When light passes through a coloured solution, some of it is absorbed, and the more absorbing molecules it meets, the less comes out. The Beer-Lambert law puts this into a formula that is the basis of spectrophotometry: absorbance is proportional to the path length, the concentration and a constant that depends on the substance and the wavelength. This calculator solves the law for whichever quantity you need and converts between absorbance and percent transmittance.
The law
The absorbance A is defined as A = −log₁₀(I/I₀) = −log₁₀(T), where T is the fraction of light transmitted. The Beer-Lambert law states A = ε · l · c, where ε is the molar absorptivity (also called the extinction coefficient) in L·mol⁻¹·cm⁻¹, l the path length through the sample in cm and c the concentration in mol/L. Absorbance has no unit. An absorbance of 1 corresponds to 10% transmittance, 2 to 1%, and 0.3 to about 50%.
Using the calculator
Under Solve for, choose Absorbance A, Concentration c, Path length l, Molar absorptivity ε, or Absorbance from % transmittance, then enter the other values. Concentrations accept scientific notation such as 5e-5. The page opens on ε = 6,220, l = 1 cm and c = 5 × 10⁻⁵ mol/L, a value typical of NADH at 340 nm, and gives A = 0.311, T = 0.48865236 (48.865236%), 51.1348% absorbed and the concentration as 0.05 mmol/L = 50 µmol/L.
Worked examples
Concentration from a reading: A = 0.45, ε = 15,000 and l = 1 cm give c = 0.45 ÷ 15,000 = 3 × 10⁻⁵ mol/L, which is 30 µmol/L, with transmittance 35.481339%. Absorbance from transmittance: 25% transmitted is A = −log₁₀ 0.25 = 0.60205999. Molar absorptivity: A = 0.52, l = 1 cm and c = 4 × 10⁻⁵ M give ε = 13,000. Path length: with ε, c and A known, l = A / (ε c).
Choosing the right range
The page warns when the absorbance exceeds about 1.5, which is only about 3% transmittance and outside the reliable range of most instruments: dilute the sample or use a shorter cell, for example 0.1 cm instead of 1 cm. For readings below 0.05 it notes that they are close to the noise of many instruments and that a longer path or more concentrated sample may be better. The ideal zone is roughly 0.1 to 1.
Calibration curves
In practice you rarely trust a single ε from a table. You prepare several standards, measure their absorbances and plot A against c; the points should fall on a straight line through the origin whose slope is ε·l. An unknown is then read off the line. This also reveals deviations from the law, such as curvature at high concentration. The serial dilution tool plans the standards, and the regression tool fits the line.
When the law fails
The Beer-Lambert law holds for dilute solutions of a single absorbing species, measured with monochromatic light. At high concentration molecules interact, the refractive index changes and the plot of A against c bends. Chemical equilibria such as dimerisation or pH-dependent colour change the absorbing species with concentration, scattering by particles adds apparent absorbance and stray light flattens high readings. Blank the instrument with the solvent and use matched cells.
Limits
The tool does arithmetic for one wavelength at a time and does not correct for a blank, stray light, scattering or overlapping absorbers. Entered values must be positive, and concentration, path length and ε must use the stated units. It does not store a calibration or fit a curve. For a mixture, the absorbances of the components add, but this page treats one absorbing species.
Related tools
- Molarity Calculator prepares the standards and converts between concentration units.
- dB to Ratio uses the same logarithmic idea for signal levels.
- Frequency to Wavelength relates the wavelength you measure at to its frequency.
- Molar Mass Calculator turns a mass concentration into a molar one.
Checked against the current version of the Beer-Lambert Law Calculator on .